Product Peformance
(1) Five modes of output:
a) Mode 1 (Direct Current Therapy Wave) has a maximum output current of 2mA with an error
of±20 %, and selectable steps of 0.02mA and 0.1mA.
b) Mode 2 (Triangular Therapy Wave) has a maximum output current of 2mA with an error of
±20%, and selectable steps of 0.02mA and 0.1mA.
c) Mode 3 (Sinusoidal Therapy Wave) has a maximum output current of 2mA with an error of
±20%, and selectable steps of 0.02mA and 0.1mA.
d)Mode 4 (Pulse Therapy Wave) has a maximum output current of 15mA with an error of ±20%,
and selectable steps of 0.5mA and 0.1mA.
e)Mode 5 (Sham Stimulation) has no output.
(2) Maximum Load Capacity: When the output maximum load is 20kΩ, the current is adjusted to the
maximum, and the output voltage is 40V with an error of ±20%.
(3) Constant Current Output Characteristics: The output current varies within an error of ±20% under
loads from 0 to 20kΩ.
(4) Incremental Adjustment Characteristics:
Modes 1/2/3: The current is modulated in two increments of 0.02mA and 0.1mA.
Mode 4: The current is modulated in two increments of 0.1mA and 0.5mA.
(5) Therapy Waveform Frequency:
Modes 2/3 have a therapy waveform frequency of 0.5Hz with an error of ±10%.
Mode 4 (Pulse Waveform) has a therapy frequency of 0.25Hz with an error of ±10%.
(6) Therapy Time Selection: Adjustable within the range of 5 to 60 minutes, with the smallest adjustable
time interval being 5 minutes.
(7)Electrode Detachment Detection: Within 10 seconds of electrode disconnection, there will be a voice
prompt indicating electrode detachment and the end of treatment.
(8)Automatic Power-off Function: The device will automatically power off after 5 minutes of inactivity.
(9)Lead Disconnection Output Protection: If the lead disconnects, the output voltage will automatically be
cut off.
(10)Low Battery Indicator: The device will provide a low battery warning when the battery level is too low.
(11)Output Control: When initiating treatment, it should always start from 0mA, otherwise, the electrical
stimulator should not have output.
(12)Open-Circuit Voltage Output: When the stimulator output is open-circuited, the measured open-circuit
voltage at the output terminals should not exceed 50V.