New in box
production 2020
more info , please call or email
Ventilation ModesVCV(A/C) PCV(A/C) PRVC SIMV(VCV) PSVSIMV(PCV) PSV SIMV(PRVC) PSV SPONT/CPAP PSVBIVENT PSV NIV/CPAP NIV-T NIV-S/TParameters• Tidal Volume: 20~2000 ml• Respiration Rate: 1~80 bpm• Tinsp: 0.2~9 s• Tslope: 0~2 s• Tpause: 0~4 s• I:E Ratio: 1:10~4:1• FiO2: 21%~100%• Trigger Sensitivity: Pressure (-20~0 cmH2O, above PEEP)Flow (0.5~20 LPM)• PEEP: 0~35 cmH2O• Psupport: 0~70 cmH2O• Pinsp: 5~70 cmH2OSpecial ProceduresApnea Ventilation Smart Suction Manual BreathInsp/ Exp Hold ETCO2 MeasurementNebulization Waveform FreezeMonitoring• Pressure Value: Ppeak, Pplat, Pmean, Pmin, PEEP• Volume / Flow Value: Vti, Vte, MV, MVspont• Time Value: ftotal,fspont, I:E• Real Time Curves: Pressure-Time, Flow-Time, Volume-Time waveformsPressure-Volume, Volume-Flow, Flow-Pressure loops• Gas Monitoring: FiO2, ETCO2• Calculated Values: Compliance(C)Resistance(R)MVleakRSBIWOBPEEPiAlarmPaw high / low MVe high / low Circuit disconnnectFiO2 high / low Inspiration / Expiratory tidal volume lowHigh Respiration Rate Apnea AC Failure Nebulizer OnLow Battery Air /O2 supply down High / Low PEEPLeakage out of range OcclusionTechnical Data• Screen: 12〞TFT color touch screen (detachable)• Supply Gas: O2, 0.28~0.6 MPa• Power Supply: AC100~240 V, 50 Hz/60 Hz• Communication Interface: RS-232 Port, Nurse call Port, Ethernet Port• Dimension (WxDxH): 322 mm x 375 mm x 366 mm (Main Unit)547 mm x 675 mm x 950 mm (Cart)• Weight: 12.5 kg (Main Unit)25 kg (Cart)